Log & Guard system

Continuous monitoring of critical quality control parameters in your IVF laboratory protects embryos and puts time on your side.

Description A comprehensive solution for 24/7 real-time monitoring, documentation & warning including sensors cover all critical parameters in the IVF lab

Webshop log in
14840/0000 Log&Guard™ Controller
  • Description: For permanent monitoring and alarming.
14840/0200 Log&Guard™ Temperature logger, starter kit
  • Description: For conventional incubators, fridges, freezers including signal converter.
14840/0201 Log&Guard™ Temperature logger, additional device
  • Description: For conventional incubators, fridges, freezers.
14840/0404 Log&Guard™ Logger for ambient temperature and humidity
  • Description: For ambient temperature and humidity.
14840/0300 Log&Guard™ CO2 sensor
  • Description: For CO2 monitoring.
14840/0310 Log&Guard™ O2 sensor
  • Description: For O2 monitoring.
14840/0406 Log&Guard™ VoC sensor
  • Description: For monitoring of VOC.
14840/0407 Replacement detector for VoC sensor
  • Description: For annual service of VOC sensor.
28032/0082 K-systems Adaptor
  • Description: Adaptor set for K-systems G210 Plus benchtop incubator, consisting of a T-fitting and two gas sensor adapters
28032/0090 K-systems Adaptor
  • Description: Adaptor-set for K-Systems G185 ST benchtop incubator, consisting of a T- fitting and two gas sensor adapters

TAKE CONTROL OF YOUR LAB AND PUT TIME ON YOUR SIDE

Complying with quality requirements and regulations in an IVF lab is crucial, but takes time and effort.

With the Log&Guard system you can replace the daily routine of manual snapshot measurements with continuous monitoring of critical parameters, presented in an easy to use web interface.

Log&Guard is a monitoring and alert system specifically designed to meet the needs of quality control in the IVF lab. All critical culture and storage parameters such as temperature, humidity, pH, CO2/O2, VOCs and power supply are continuously measured and documented through a network of external sensors.

Start small or go big

The flexible Log&Guard system lets you decide which and how many external sensors to be connected to the main controller. You also have a choice of data transmission by cable, wirelessly or a combination of the two.

Monitor critical parameters to create a protective environment

Controller for monitoring and warning

The controller includes; integrated PC, GSM modem, battery backup, external keyboard and flat screen, 8 alarm relay inputs; LAN interface, slot for SIM card (to be provided by the user for activating the text message alert). 

There are interfaces for temperature-, CO2- , O2 sensors and pH Online

Technical data: power supply 110-240 V AC via external power adapter, external dimensions 260 x 300 x 90 mm (LxWxH), LCD display, battery service life approx. 20 min

Easy to use software program

Log&Guard operates on a web interface based software platform, easily accessible from any computer in your local network, and does not require a separate computer.

Alarm conditions and thresholds, measurement intervals and many other parameters can be set individually to perfectly adapt the monitoring network to your individual lab. Measured data can be downloaded for long term storage.

Never miss an alarm

The Log&Guard system will alert you in case any parameter is out of range. Besides acoustical and optical alerts in the lab you can also be notified by text messages on your phone.

CO2 sensor

  • Dual wavelength infrared sensor, measuring range 0 - 10% CO₂
  • Accuracy ± 5% of the indicated value
  • Resolution 0.1% 
  • Sensor diameter 43 mm, length 91 mm

O2 sensor

  • For monitoring reduced oxygen atmosphere
  • Measuring range 0 - 25% O₂
  • Accuracy ± 0.5%
  • Resolution 0.1%.
  • Sensor diameter 43 mm, length 44 mm

VOC sensor

  • Detects embryotoxic volatile organic compounds (VOC) in ambient air
  • Super sensitive measuring range 0 - 20 ppm (parts per million)
  • Accuracy ± 0.1 ppm
  • Resolution 0.01 ppm
  • Reaction time < 4 sec

Temperature logger 

  • Starter kit for conventional incubators, fridges, freezers
  • Pt1000 sensor
  • Temperature range -50 to +150°C
  • Accuracy ± 0.5°C
  • Resolution 0.1°C
  • Battery backed (12 years battery life)
  • Level converter supporting up to nine loggers
  • Dimensions (housing) 48.5 x 48.5 x 35.5 mm (LxWxD)

Logger for ambient temperature and humidity

  • Pt1000 sensor combined with a capacitive polymer humidity sensor
  • Temperature range -25 to + 60°C
  • Accuracy ± 0.5°C
  • Resolution 0.1°C
  • Humidity range 0 to 100% RH
  • Accuracy ± 3% RH
  • Battery backed (6 years battery life)
  • Dimensions (housing) 48.5 x 48.5 x 35.5 mm (LxWxD)